Model No.︰GF1061
Brand Name︰GFUVE
Country of Origin︰China
Unit Price︰US $ 16000 / pc
Minimum Order︰1 pc
Share on:

 Total 19 Related Items 
Prev45678Next

Product Description

Detailed Product Description

Name:

CT PT Analyzer

Application:

Current Transformer And Potential Transformer Test

Theory:

DC Variable-frequency Methods

LCD:

7inch TFT Touch Color LCD

Phase Error:

1min Or 3min

Ratio Error:

0.02% Or 0.05%

CT Raito:

1-50000

PT Ratio:

1-10000

Usage:

Measurement CT And Protection CT

Weight

4KG

 

GF1061 portable CT PT analyzer is mainly used for field or lab testing, it can finish the measurements (M) and protection (P) class CT, PT and TYP class CT. Adopt 7 inch touch TFT LCD, self-equipped mini type printer supporting field printing; supporting to use USB flash disk to dump data. The CT PT Analyzer is the most complete and easy-to-use testing system for protection and metering CTs according to IEEE and IEC standards.

The CT Analyzer automatically determines all relevant current transformer values and compares the results with the selected standard. The device evaluates the current transformer as per the IEC or IEEE standard within seconds with the push of a button. its can test current transformer and potential transformer ratio error and phase error, Magnetization curve etc
It facilitates fast and economical on-site testing and calibration of metering and protection current transformers. Manufacturers of current transformers, high voltage transformers and switchgear also use the CT Analyzer for tests during development and production.

Application

1. Measurement type current transformer

2. Protection type current transformer

3. GIS type current transformer

4. Bushing type current transformer

5. Zero sequence current transformer

6. Voltage transformer

MAIN FUNCTIONS

I. Current Transformer (CT)

II. Voltage Transformer (PT)

1. Magnetization curve

1. Excitation characteristic test

2. Transformation ratio test

2. Transformation ratio test

3. Polarity

3. Polarity

4. 5% and 10% error curve

4. Ratio error, phase error

5. Accuracy limiting factor (ALF)

5. Degauss

6. Degauss

6. Calculation of knee point value

7. Ratio error, phase error

7. Actual secondary load test (burden test)

8. Automatic calculation of excitation knee
point value

8. Resistance test

9. Actual secondary load test (burden test)

10. Resistance test 

11. Secondary time constant (Ts)

12. Remanence coefficient (Kr)

13. Transient dimensioning factor (Ktd)

14. Peak instantaneous error (Er)

15. Magnetizing inductance (LU)

16. Instruments security factor(FS)

17. Composite error

 

 

PARAMETERS

 

 

Electrical parameters

Accuracy

0.05%, 0.02%

Power supply

AC 220V±10% or AC 120V±10%, 50/60Hz or Battery

Output voltage

0-100Vrms

Output current

0-5Arms (20A peak-value)

Output power

0-400 VA (1500 VApeak)

Automatic frequency variation range

0.1-60Hz

Equivalent excitation voltage

≤5000V

Accuracy

≤0.5%

Secondary winding DC resistance measurement

Range

0.1-1000Ω

Accuracy

≤0.05%

Secondary actual load measurement

Range

5VA-1000VA

Accuracy

≤0.5%±0.1VA

CT/PT phase error measurement

Accuracy

±1min (typical) / 3 min (guaranteed)

Resolution

0.1min

CT ratio error measurement

Range

1-50000

Accuracy

≤0.05%

PT ratio error measurement

Range

1-10000

Accuracy

≤0.05%

Standards 

Reference standards

GB1207-2006, GB1208-2006, GB16847-1997
IEC60044-1, IEC60044-6, IEC61869-2, ANSI/IEEE C57.13

Safety standards

GB 4793.1-2007

EMC

EMC standard 89/336/EEC

FCC Subpart B of Part 15 Class A

IEC 1000-4-2 /3 /4 /6

Mechanical parameters 

Overall dimension (L x W x H) (mm)

280 x 250 x 160

Weight (kg)

≤4

Environmental conditions 

Relative humidity

Relative humidity 5%-95% not condensing

Operating temperature

-10°C to +50°C

Storage temperature

-20°C to +70°C

Altitude

≤1000m

 

 

 

 

Price Terms︰FOBBeijing EXWBeijing CPT
Payment Terms︰TT/Western Union
Product Image






Related Products
AC Current Probe for cable fault tester
AC Current Probe for cable fault tester
US $ 150
Current Clamp Probe
Current Clamp Probe
US $ 150
Zero Sequence Current Transformer
Zero Sequence Current Transformer
US $ 120
Medium Voltage Split Core Current Transformer
Medium Voltage Split Core Current Transformer
US $ 120




Home  |  Product Catalog  |  News  |  Contact Us  |  Sitemap  |  Mobile Version

Powered by DIYTrade.com  Build your FREE website !
Find Best Price For China Products !